The following table describes the tabs in the Instrument Control pane.
No.
Tab
1
Gain
2
Threshold
3
Signal
4
Lasers
Worksheet Area
The worksheet area allows you to view the data in plots and create plots, statistics, and gates.
Description
Gains can be adjusted for all detector channels for all lasers using the
gain spinboxes. FSC gain can be adjusted from 0–1,000. SSC and
fluorescence detector gains can be adjusted from 0–10,000. To change
the value that the gain increments, see the acquisition preferences on
page 59.
Use the Threshold tab to set the threshold parameter and minimum
threshold channel value. Multiple parameters can be set as a
threshold using either the AND or OR operator. Use OR for at least one
parameter to be available.
Use the Signal tab to select area, height, or width for each signal. Area
and height can be selected for all channels. Width can be selected for
only one channel per laser.
Use the Lasers tab to set the area scaling factor and laser delay. These
values are automatically set and updated in all user settings upon
completion of the Daily QC.