Low Priority Alarm

A low priority alarm indicates that an equipment fault has occurred and the device is unable to
provide a measurement value. See table 8 for parameters, fault messages, and possible cause.
Low priority alarms are both audible and visual:
• Audible alarms beep slower in a low priority situation than in a high priority situation.
• The fault message displays on the monitor.
Table 8: Low Priority Alarm Parameters and Causes
Parameter
Message
Pulse oximetry
NO PROBE
Pulse oximetry
NO FINGER
Pulse oximetry
ARTIFACT
Pulse oximetry
SIGNAL LOW
Capnometry
OCCLUSION*
Capnometry
TRAP FULL?
PUSH ALARM
System
BATT LOW
System
DISP ERROR
System
NO OXIMET
System
NO CAPNO
*A full moisture trap or a kinked sampling line or cannula may trigger the occlusion alarm. To
prevent the monitor from damage by liquid, the pump will stop after 10 seconds of occlusion and
the message "TRAP FULL? / PUSH ALARM" displays. Check the moisture trap and replace it if
necessary. Check the sampling line or cannula for kinks or occlusions and replace if necessary.
Press the Audible Alarm Pause/Resume

Disable Alarms

It is possible to disable the audible alarms either by selecting Audible Alarm Disable
start-up screen or by decreasing all lower limit settings to 0. When audible alarms are disabled,
the Alarm Disabled icon displays on the touch panel display.
Possible Cause
Sensor is not connected to the monitor.
Sensor is not connected to the finger.
A questionable pulse was detected
Hard to detect a pulse. Verify perfusion status at the sensor
application site, minimize motion, and verify that there is not
excessive ambient light.
Low or no flow from sample line tubing or cannula.
There has been an occlusion for several seconds, possibly
due to moisture in the moisture trap. Replace it and then
press the Audible Alarm Pause/Resume button.
Battery is almost depleted.
Touch panel display is not working properly.
No communication from the pulse oximetry unit.
No communication from capnometry unit.
button to continue.
on the
27