Low Priority Alarm; Disable Alarms; Table 7. Low Priority Alarm Parameters And Causes - Nonin RespSense LS1R-9R Operator's Manual

Capnography monitor
Table of Contents

Advertisement

Low Priority Alarm

A low priority alarm indicates that an equipment fault has occurred and the device is unable to
provide a measurement value. See table 7 for parameters, fault messages, and possible causes.
Low priority alarms are both audible and visual:
• Audible alarms beep slower in a low priority situation than in a high priority situation.
• The fault message displays on the monitor.
Table 7: Low Priority Alarm Parameters and Causes
Parameter
Message
Capnometry
OCCLUSION* Low or no flow from sample line tubing or cannula.
Capnometry
TRAP FULL?
PUSH ALARM
Capnometry
NO CAPNO
Capnometry
WARM UP
System
BATT LOW
System
DISP ERROR
*A full moisture trap or a kinked sampling line or cannula may trigger the occlusion alarm. To
prevent the monitor from damage by liquid, the pump will stop after 10 seconds of occlusion and
the message "TRAP FULL? / PUSH ALARM" displays. Check the moisture trap and replace it if
necessary. Check the sampling line or cannula for kinks or occlusions and replace if necessary.
Press the Audible Alarm Pause/Resume

Disable Alarms

It is possible to disable the audible alarms either by selecting Audible Alarm Disable
start-up screen or by decreasing all lower limit settings to 0. When audible alarms are disabled,
the Alarm Disabled icon displays on the touch panel display.
22
There has been an occlusion for several seconds, possibly
due to moisture in the moisture trap. Replace it and then
press the Audible Alarm Pause/Resume button.
No communication from capnometry unit.
Warm up delay and stabilizing measurements.
Battery is almost depleted.
Touch panel display is not working properly.
button to continue.
Possible Cause
on the

Advertisement

Table of Contents
loading

Table of Contents