Capnography Specifications - Nonin LifeSense LS1-9R Operator's Manual

Vital signs monitor capnography/pulse oximeter
Table of Contents

Advertisement

Capnography Specifications

Respiration Range:
Update Frequency:
Respiration Accuracy:
EtCO
/CO
Range:
2
2
EtCO
/CO
Accuracy:
2
2
Update Frequency:
Sampling Rate:
Total System Response Time:
Drift of Measurement:
Measurement:
Presented concentration of CO
interfering gases.
The table below shows the CO
the table below.
Agent Concentration
50 – 70% N
O
2
30 – 50% N
O
2
0 – 30% N
O
2
0 – 5% Isoflurane
0 – 4% Halothane
40
3 to 60 respirations/minute
Once every breath (No Breath alarm after 25 seconds)
3 to 50 respirations/minute ±2
51 to 60 respirations/minute ±3
0 to 9.9 kPa or 0 to 99 mmHg
±0.2 kPa / ±2 mmHg +8% of reading†
540 – 795 mmHg
(EtCO
/CO
reading reaches its steady state accuracy 10 minutes
2
2
after power up)
Once every breath (No Breath alarm after 25 seconds)
4 Hz (4 times per second)
4 seconds (includes delay time and rise time)
Within CO
accuracy specifications for 6 hours of continuous
2
monitoring
Automatic barometric pressure compensation and CO
compensation
and EtCO
can be false, indicating a high presence of nitrous oxide and other
2
2
and EtCO
concentration corrections. Only use agents listed in
2
2
Correction of Presented CO
Presented CO
x 0.75 = Actual CO
2
Presented CO
x 0.85 = Actual CO
2
No correction
No correction
Presented CO
x 0.98 = Actual CO
2
to Real Concentration
2
2
2
2
temperature
2