Nafion Tubing; Attaching The Nafion Tubing; Single-Patient Use, Disposable Moisture Trap And Filters; Replacing The Moisture Trap/Filter - Nonin LifeSense LS1-9R Operator's Manual

Vital signs monitor capnography/pulse oximeter
Table of Contents

Advertisement

Nafion Tubing

The Nafion tubing is a single-use disposable component designed to be placed between the
moisture trap and the nasal cannula or sampling tubing to remove water vapor. It is intended for
use only with Nonin's LifeSense and RespSense monitors.
CAUTION: Water or other liquid in the sampling tube may cause erroneous CO
!
readings.
CAUTION: Ensure that all connections are tight, leak-free, and properly attached.
!
CAUTION: If the Nafion tubing becomes contaminated or damaged during use,
!
discard it and replace it with a new one.

Attaching the Nafion Tubing

1. Connect male end of the Nafion tubing to the moisture trap. Turn clockwise to tighten.
2. Connect female end of the Nafion tubing to the sampling line or cannula. Turn clockwise
to tighten.
3. Ensure that the Nafion tubing is firmly attached.
Single-Patient Use, Disposable Moisture Trap and
Filters
The moisture trap and filters are single-use disposable components. During long-term monitoring
of a patient, the moisture trap fills up with liquid (condensed moisture from breathing). Check the
moisture trap frequently and replace when necessary.
Make sure to keep a sufficient supply of new moisture traps and filters within easy reach.
When the moisture trap is removed, guide marks (numbered 1 and 2) and arrows, are visible on
the back of the monitor. These guide marks help the operator insert the moisture trap.

Replacing the Moisture Trap/Filter

CAUTION: The sample line, moisture trap, Nafion tubing, filter, and T-connector are
!
single-patient use, disposable components. Do not disassemble the plastic parts of
the single-use disposable moisture trap. Dispose all components in accordance with
your local, state, or national regulations regarding waste management.
1. Place the filter in the moisture trap as shown in figure 6 (1).
2. Slide the moisture trap into position (figure 6, 2) using the guide marks on the back of the
monitor.
3. Press the moisture trap into position using the tab (figure 6, 3).