Battery Care; Maintenance; Ensuring Optimal Performance; Cleaning The Sensor - Nonin LifeSense LS1-9R Operator's Manual

Vital signs monitor capnography/pulse oximeter
Table of Contents

Advertisement

Battery Care

The battery, made of Lithium Ion (Li-Ion) rechargeable cells, is integral to the device and cannot
be replaced by anyone other Nonin Technical Service. The life expectancy of the battery is
approximately 1 year.
For optimal performance, the battery should be replaced once per year to limit the amount of Li
build up if the battery is charged in a cold environment.

Maintenance

Ensuring Optimal Performance

In order to ensure safety and optimal performance of LifeSense, Nonin recommends a yearly
inspection and functional check be performed on the monitor (see Recommended Inspections
and Functional Check section). This inspection and functional check may be performed by Nonin
Technical Services or at your facility. Additionally, the LifeSense monitor should be calibrated
(see Calibration section), and the calibration should be verified using 5% CO
apparatus, gas valve, and 5% CO
Please contact Nonin Technical Services if monitor maintenance cannot be performed at your
facility.
CAUTION: Always turn off the monitor prior to cleaning the monitor or changing the
!
pulse oximeter sensor or moisture trap and/or filter.

Cleaning the Sensor

Refer to individual sensor Instructions for Use for details.

Cleaning the Monitor

Clean the monitor with a soft cloth moistened with isopropyl alcohol. Allow the monitor to dry
completely after cleaning.
CAUTION: Do not sterilize or autoclave the monitor or sensors. Do not immerse in
!
liquids.

Calibration

LifeSense has a built-in zero-point calibration function for CO
at least every 6 months, or if the baseline of the CO
The calibration apparatus (see "Accessories") is reusable for approximately 100 times. When the
pellets start to turn purple they cannot absorb any more CO
be replaced. Dispose of the calibration apparatus in accordance with your local, state, or national
regulations concerning waste materials.
verification gas are available from Nonin [see Accessories]).
2
graph is elevated.
2
gas (a calibration
2
. Perform the calibration procedure
2
and the calibration apparatus must
2
29