4. Verify the moisture trap and filter are in position.
5. Verify the reusable PureLight finger clip sensor is clean, if previously used. Visually
examine the accessories for defects prior to use.
6. Turn on the monitor by pressing the ON/OFF
7. Verify all parameters display correctly and adjust any alarm limits according to the patient.
8. Verify alarm function/status by simulating alarm situations for all parameters.
9. Visually verify the zero-point of the CO
WARNING: If the LifeSense fails to respond as described, discontinue use and
contact Nonin Technical Service.
CAUTION: LifeSense should only be operated by trained licensed practitioners.
!
WARNING: Never allow liquids to enter into or to be spilled onto the monitor. If liquid
has penetrated into the monitor it must be checked by Nonin Technical Service.
CAUTION: Be careful not to drop LifeSense on the floor or strike it against hard
!
surfaces. If such an incident happens, do not use LifeSense until a functional test
has been carried out.
CAUTION: Never open the monitor housing/case. By opening the case you render
!
your warranty invalid.
button until you hear a beep.
graph is not elevated.