System Specifications - Nonin LifeSense LS1-9R Operator's Manual

Vital signs monitor capnography/pulse oximeter
Table of Contents

Advertisement

System Specifications

Power Data
Power Consumption: 3.6 W with battery operation
Battery Data
Physical Data
Operation
Working Temperature: 23 °F to 104 °F (-5 °C to 40 °C)
Device temperature will not exceed 41°C as measured during a controlled environment test.
Atmospheric Pressure: 720 to 1060 hPa (540 to 795 mmHg)
Storage
Pump
Alarms
Sound Pressure Level: 65 dBa maximum at 1 meter in front of monitor
Classification per IEC 60601-1 / CAN/CSA-C22.2 No. 601.1 / UL 60601-1:
Type of Protection: Internally powered class II (with battery charger)
Degree of Protection: Type BF-Applied Part
Mode of Operation: Continuous
Enclosure Degree of Ingress Protection: IPX2
38
Power Supply: 100 – 240 VAC 50 – 60 Hz
9 W with power supply
Input: 12 VDC, 720 mA
Type: Lithium Ion (Li-Ion) internal battery, non-field replaceable,
rechargeable
Battery Capacity: Approximately 8 hours
Charging Time: Approximately 17 hours, or 2 hours for each hour of use
Dimensions: 200 x 135 x 50 mm (7.9 x 5.3 x 2 in.)
Weight: 800 grams (1.8 pounds)
Humidity: 10 % to 90 % (non-condensing)
Altitude: Up to 9,000 ft (2,740 m)
Temperature: -40 °F to 158 °F (-40 °C to 70 °C)
Humidity: 10 % to 90 % (non-condensing)
Pump Flow: 75 ml/min
Flow Accuracy: ±15 ml/min