To Connect DUT
Set the DUT to the test xture. For details, see \Test Fixtures and Test Leads".
Do
Warning
not
when the High Voltage indicator is ON, the 4339B outputs dangerous
voltage of up to 1000 Vdc. Before handling the 4339B or the accessory,
turn OFF the test voltage pressing
indicator is OFF.
To Select Measurement Parameter
1. Press
. The measurement parameter selection menu will be displayed.
2. Select the desired parameter using
:Volume resistivity
Rv
If the 16008B resistivity cell is connected to the 4339B, to change volume and
Note
surface resistivity, switch the volume/surface selector on the resistivity cell.
2-6
Operating the 4339B
touch the UNKNOWN terminals or the electrodes of the accessory,
or
v,
:Surface resistivity
Rs
and conrm that the High Voltage
and press
. (
R
s)
4339B
:Resistance,
:Current,
I