The following settings will increase the dynamic range of your NA mode
measurement.
 Increase Power Level: Press Meas 4 then
Output Power High
 Lower the IFBW: Press BW 2 then
IF BW
 Increase Averaging: Press BW 2 then
Average
The following procedure MAY increase the dynamic range of your NA mode
measurement. The results you see will depend on the performance of your DUT.
With an S21 trace active:
1. With RF OUT (port-2) open, press Trace 6 then
Math and Memory
then
Data->Mem
2. Re-connect the DUT.
Data Math
Data-Mem
3. Press
then
52