Resolution (Number Of Points) - Agilent Technologies FieldFox N9913A User Manual

Hide thumbs Also See for FieldFox N9913A:
Table of Contents

Advertisement

How to set Smoothing
 Press BW 2.
 Then
 Then
 Press Enter
Single or Continuous Measure
This setting determines whether the FieldFox sweeps continuously or only once
each time the
to allow you to save or analyze a specific measurement trace.
This setting can be changed at any time without affecting calibration accuracy.
How to set Single or Continuous
 Press Sweep 3.
 Then choose one of the following:
You can also use Run / Hold +/- to toggle between Single and Continuous.
Resolution (Number of Data Points)
Data points are individual measurements that are made and plotted across the X-
axis to form a trace. Select more data points to increase measurement resolution.
However, more data points also takes more time to complete an entire
measurement sweep.
When the Resolution is changed after a calibration is performed, the cal becomes
interpolated. Learn more on page 71.
How to set Resolution
 Press Sweep 3 .
 Then Resolution.
 Then choose from the following:
 Using SCPI, Resolution can be set to ANY number of points between 3 and
NA (Network Analyzer) Mode
Smoothing ON OFF
Sm. Aperture
and enter a value between 0 and 25 (percent) using the
numeric keypad.
Single
button is pressed. Use Single to conserve battery power or
Single
Automatically sets Continuous OFF and causes FieldFox to make
o
ONE sweep, then hold for the next Single key press.
the upper left corner of the display when NOT sweeping, and changes to an
arrow
-->
while the sweep occurs.
Continuous
Makes continuous sweeps. This is the typical setting when
o
battery power is not critical.
101 |201 | 401 | 601 | 801 | 1001 | 1601 | 4001 |
10001. See the Programming Guide at
Hold
is annotated in
10001.
http://na.tm.agilent.com/fieldfox/help/
47