Center Freq = 1.80 GHz
The Loss value for the measurement is interpolated from the Freq/Loss pairs at
the Center Freq:
1 GHz = 0.1 dB/m
1.8 GHz = 0.18 dB/m
2 GHz = 0.2 dB/m
The correction for loss at 5 meters in one direction: 0.18 dB/m * 5m = 0.9 dB.
All DTF measurements correct for loss for travel down the DUT and back, so
double the correction: 0.9 dB * 2 = 1.8 dB.
Window Settings
Window settings provide the ability to choose between optimizing DTF
measurements for resolving closely-spaced faults or for the ability to measure
low-level faults.
How to select Window settings
 Press Meas Setup 4.
 Then Settings.
 Then press ▲|▼ arrows to move to the Window row.
 Then press Edit.
 Then press
 Then press
 Again press Done.
DTF Units
The DTF Units setting is available ONLY on the DTF Settings table.
By default, X-axis units for DTF measurement settings are displayed in Meters.
How to change DTF units
 With a DTF measurement present, press Freq/Dist .
 Then
 The current selection is underlined
Calculated DTF values
Press
Values noted on the FieldFox screen with c - <setting>
Start Frequency – Start frequency that is used to calculate DTF.
36
Window
repeatedly and choose from the following:
Maximum – Optimized for dynamic range, the noise floor is lowered to
o
provide the ability to measure low-level responses. (Default setting)
Medium – Compromise between Min and Max window settings.
o
Minimum – Best Response Resolution, providing the ability to resolve
o
between two closely-spaced responses.
Done
Edit.
DTF
Units.
Next Page
on the DTF Settings Table to view the following calculated
m (meters)
Feet.
FieldFox User's Guide