The voltage is output through an SMB(m) output connector located on the left-
side panel. See the connector on page 17.
An optional SMB (f) to BNC (m) bias-tee power cable is available with Agilent
part number: N9910X-713.
 The following status line is provided near the top of the FieldFox screen.
Voltage Source Status Line
1.
2.
3.
4.
How to make Voltage Source settings
 Press System 7 .
 Then
 Then
 Then
 Then

Security Level

For security reasons, you can prevent frequency information from appearing on
the FieldFox while in SA, NA, or CAT modes.
How to set Security Level
 Press System 7
 Then
 Then
 Then choose from the following:
System Settings
Voltage source state: ON, OFF, TRIPPED.
Voltage setting from 1.00 V to 32.00 V with .01 V resolution.
Current: The amount of current (in Amperes) being drawn from the source.
Max Current: A conservative estimate of maximum current possible given
the present voltage setting. The actual 'trip' level may be slightly different.
System
Configuration.
Voltage
Source.
Source ON OFF
to enable and disable the voltage source output.
Voltage x.xx V
Using ▲|▼ arrows, the rotary knob, or numeric keypad,
enter a value between .001 and 32 volts.
Clear Overload
If the voltage source becomes 'TRIPPED' press to re-enable
the voltage source output. Then reset
Status Line ON OFF
Turn ON and OFF the voltage source status line that
appears near the top of the screen. The status line is always ON while making
Voltage Source settings.
System Configuration
Security Level
None
All frequency settings are visible.
o
High
Frequency information is blanked from the following:
o
o
Display annotation
o
Softkeys
o
Marker display and marker table
Source ON.
183