Application; Indications For Use; Conditions Of Use; Indications - Otto Bock C-Leg 3C98-3 Instructions For Use Manual

Hide thumbs Also See for C-Leg 3C98-3:
Table of Contents

Advertisement

Please ensure that the qualified personnel applying the product is not only authorised for fitting this knee joint, but
also for the connection to the osseointegrated implant system.

3 Application

3.1 Indications for use

The product is to be used solely for lower limb exoprosthetic fittings.

3.2 Conditions of use

The product was developed for everyday use and must not be used for unusual activities. These unusual activities
include, for example, extreme sports (free climbing, parachuting, paragliding, etc.).
Permissible ambient conditions are described in the technical data (see Page 39).
The product is intended exclusively for use on one patient. Use of the product by another person is not approved
by the manufacturer.
Our components perform optimally when paired with appropriate components based upon weight and mobility
grades identifiable by our MOBIS classification information and which have appropriate modular connectors.
The product is recommended for mobility grade 2 (restricted outdoor walker), mobility grade 3
(unrestricted outdoor walker) and mobility grade 4 (unrestricted outdoor walker with particularly
high demands). Approved for a body weight of 136 kg max.
kg

3.3 Indications

For patients with knee disarticulation, transfemoral amputation and hip disarticulation
For unilateral or bilateral amputation
Dysmelia patients with residual limb characteristics corresponding to knee disarticulation, transfemoral ampu­
tation or hip disarticulation
The patient must fulfil the physical and mental requirements for perceiving visual/acoustic signals and/or mech­
anical vibrations.

3.4 Qualification

The product may be fitted only by qualified personnel authorised by Ottobock after completing the corresponding
training.
If the product is to be connected to an osseointegrated implant system, the qualified personnel must also be
authorised for the connection to the osseointegrated implant system.