Additional operating states (modes)

3x – 5x
1s
ð
1) Position the prosthetic leg back slightly.
2) While maintaining constant contact with the floor, bounce on the forefoot at least 3 times but not more than
5 times.
3) Keep the prosthetic leg still in this position (lunge position) for about 1 second without lifting the leg. Taking the
weight off is no longer necessary.
→ A beep and vibration signal will occur to confirm that the movement pattern has been recognised.
INFORMATION: If this beep and vibration signal is not emitted, the requirements were not met
while bouncing.
4) Move the prosthetic leg in next to the contralateral leg, set it down and keep it still for approx. 1 second.
→ A confirmation signal will sound to indicate that the prosthesis has successfully switched over to basic mode.
INFORMATION: If this confirmation signal does not sound, the leg with the prosthesis was not cor­
rectly repositioned and kept still. Repeat the process to correctly switch to the required mode.
11 Additional operating states (modes)

11.1 Empty battery mode

Beeps and vibration signals are emitted if the available battery charge level is 0% (see Page 42). During this time,
damping settings are set to their safety mode values. The prosthesis is then switched off. You can switch back to
basic mode (mode 1) from empty battery mode by charging the product.

11.2 Mode for charging the prosthesis

The product is non-functional during charging.
The product is set to safety mode damping. This may be low or high depending on the setting in the adjustment
software.

11.3 Safety mode

The product automatically switches to safety mode if a critical system fault occurs (e.g. failure of a sensor signal).
Safety mode remains in effect until the error has been rectified.
Default damping values are activated in safety mode. This makes limited walking possible for the user even though
the system is not active.
The switch to safety mode is indicated by beeps and vibration signals immediately prior to switching (see Page 42).
Safety mode can be disabled by connecting then disconnecting the battery charger. If the product switches into
safety mode again, this means a permanent error exists. The product must be inspected by an authorised Ottobock
Service Centre.

11.4 Overheating mode

When the hydraulic unit overheats due to uninterrupted, increased activity (e.g. extended walking downhill), damp­
ing is increased along with the rising temperature in order to counteract the overheating. When the hydraulic unit
cools down, the product switches back to the damping settings that existed before the overheating mode.
Overheating mode is not activated in the MyModes.
Overheating mode is indicated by a long vibration every 5 seconds.
The following functions are deactivated in overheating mode:
Sitting function
Display of the battery charge level without additional equipment
Switching to a MyMode
Changes to the prosthesis setting
36
1x
1x
ð
+
1s
ð
ð
1x
1x
ð
Basic mode
+
C-Leg 3C98-3/3C88-3
þ

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

C-leg 3c88-3

Table of Contents