Beckman Coulter Optima XE Series User Manual page 98

Preparative ultracentrifuges for non-in vitro diagnostic use
Table of Contents

Advertisement

Diagnostics
Diagnostics/User Messages Chart
Table C.1 Diagnostics/User Messages Chart (Continued)
Message
D211 - Power Supply
D212 - Power Supply
D213 - Power Supply
D214 - Power Supply
D216 - AC Power Out of
Range
A217 - Alert: AC Power
Voltage Sag
D300 - Rotor Speed
A301 - Alert: Rotor
Speed Adjusted
A302 - Alert: Inertia
Calibration
D303 - Speed Signals
D304 - Speed Signals
C-4
Definition/Result
SCB +3.3V supply too high
(>3.63VDC)
SCB analog +3.3V supply
too high (>3.63VDC) or
SCB analog +3.3V supply
too low (<2.97VDC)
System +24V supply too
high (>26.4VDC) or
System +24V supply too
low (<21.6VDC)
System +5V supply too
high (>5.5VDC)
AC Power Out of Range
AC Power Voltage Sag
below 180Vac
Rotor speed exceeds
maximum rated speed
Rotor set speed adjusted
Inertia calibration failure
Overspeed timing signal is
<9 counts/revolution or
>47 counts/revolution, or
overspeed timing signal is
unstable
No tachometer signal
Recommended Action
Call Beckman Coulter Field Service.
Call Beckman Coulter Field Service.
Call Beckman Coulter Field Service.
Call Beckman Coulter Field Service.
Check electrical power source.
None
Call Beckman Coulter Field Service.
None
Call Beckman Coulter Field Service.
1. Check the condition of the rotor overspeed
disk.
2. Make sure the rotor is installed properly.
3. Verify the rotor load is within limits specified
in the rotor manual.
NOTE
This condition will result in a 20-minute/
10K of RPM delay before the diagnostic can
be cleared and the door opened. The power
must be left on, until the delay period is
completed.
1. Make sure the rotor is installed properly.
2. Check the condition of the rotor overspeed
disk.
3. Verify the rotor load is within limits specified
in the rotor manual.
4. If the problem persists, call Beckman Coulter
Field Service.
PN B07711AA