Physical Data - Beckman Coulter Optima XE Series User Manual

Preparative ultracentrifuges for non-in vitro diagnostic use
Table of Contents

Advertisement

Specifications
Temperature
Ease of Use

Physical Data

Specification
Dimensions
Ventilation Clearances
PN B07711AA
Description
• Set temperature:
— 0 to 40°C in 1°C increments
• Rotor Temperature (after equilibration):
— ± 0.5°C over entire set temperature range for all rotors except
SW-32
— ± 0.8°C over entire set temperature range for SW-32
• Temperature display:
— Actual rotor temperature in 0.1°C increments
• Ambient temperature range:
— 10 to 35°C
• Humidity restrictions:
— <80% at <35°C (non-condensing)
• Condensation removal:
— Automatic (begins five minutes after vacuum vent)
• Vacuum:
— Below 5 microns (0.7 Pa)
• Languages:
— Multi-language support
• Help:
— On-screen, context sensitive
• Diagnostic messages:
— 10,000 messages saved
Description
• Width :
— 94.0 cm (37 in.)
• Depth:
— 68.1 cm (26.8 in.)
• Height:
— 125.7 cm (49.5 in.)
• Weight
— 485 kg (1068 lb)
• Sides
— 5.1 cm (2.0 in.)
• Rear
— 15.2 cm (6.0 in.)