Reference Channel - Ametek 7230 Instruction Manual

Dsp lock-in amplifier
Table of Contents

Advertisement

maximize the X channel output and minimize the Y channel output.
ACGAIN [n]
AC Gain control
Sets the gain of the signal channel amplifier according to the following table:-
n
AC Gain
0
0 dB
1
6 dB
2
12 dB
3
18 dB
4
24 dB
5
30 dB
6
36 dB
7
42 dB
AUTOMATIC [n] AC Gain automatic control
n
Status
0
AC Gain is under manual control, either using the Main Controls panel or the
ACGAIN command
1
Automatic AC Gain control is activated, with the gain being adjusted according
to the full-scale sensitivity setting
LF [n
n
]
Signal channel line frequency rejection filter control
1
2
The LF command sets the mode and frequency of the line frequency rejection
(notch) filter according to the following tables:
n
Selection
1
0
Off
1
Enable 50 or 60 Hz notch filter
2
Enable 100 or 120 Hz notch filter
3
Enable both filters
n
Notch Filter Center Frequencies
2
0
60 Hz (and/or 120 Hz)
1
50 Hz (and/or 100 Hz)

6.7.02 Reference Channel

REFMODE [n]
Reference mode selector
The value of n sets the reference mode of the instrument according to the following
table:
n
Mode
0
Single Reference / Virtual Reference mode
1
Dual Harmonic mode
2
Dual Reference mode
NOTE: When in either of the dual reference modes the command set changes to
accommodate the additional controls. These changes are detailed in section 6.7.14
IE [n]
Reference channel source control (Internal/External)
In Single Reference and Dual Harmonic mode, the value of n sets the reference input
mode according to the following table:
Chapter 6, COMPUTER OPERATION
n
AC Gain
8
48 dB
9
54 dB
10
60 dB
11
66 dB
12
72 dB
13
78 dB
14
84 dB
15
90 dB