Ametek 7230 Instruction Manual page 17

Dsp lock-in amplifier
Table of Contents

Advertisement

Figure 2-3, Model 7230 Main Controls Panel
9) Click the Equations tab, to show the Equations panel, as shown below in figure
2-4
10) Click the Auto Default button. This will set all instrument controls to the factory
default values,
11) Connect a BNC cable from the front panel OSC OUT to the A input connector
on the front panel. The X1 and Magnitude outputs should read 100 ± 1%, and the
Y1 and Phase 1 outputs should read close to zero, as shown in figure 2-5 below.
Chapter 2, INSTALLATION AND INITIAL CHECKS
Figure 2-4, Model 7230 Equations Panel
2-3

Hide quick links: