Ametek 7230 Instruction Manual page 163

Dsp lock-in amplifier
Table of Contents

Advertisement

n
Trigger
0
Trigger generated once per curve, at start of curve
1
Trigger generated once per point
TRIGOUTPOL [n]
The value of n defines the polarity of the trigger output generated when a buffer
acquisition is run in accordance with the following table:
n
Trigger Output
0
Rising-edge
1
Falling edge
USBTERM [n]
Set/read USB terminator
The value of n sets the USB terminator according to the following table:
n
Function
0
A null character (ASCII 0)
1
A null character (ASCII 0), followed by a byte representing the value of the
status Byte (table 6-1), followed by a byte representing the value of the Overload
Byte (table 6-1)
VER
Report firmware version
Causes the lock-in amplifier to respond with the firmware version number, as also
displayed on the Configuration menu.
VMODE [n]
Voltage input configuration
The value of n sets up the input configuration according to the following table:
n
Input configuration
0
Both inputs grounded (test mode)
1
A input only
2
-B input only
3
A-B differential mode
Note that the IMODE command takes precedence over the VMODE command.
VRLOCK [n]
Virtual reference mode lock
The Seek option of the frequency sweep mode must be used before issuing this
command, for which the value of n has the following significance:
n
Mode
0
Disables virtual reference mode
1
Enters virtual reference mode by enabling tracking of the signal frequency
X[.]
X channel output
In fixed point mode causes the lock-in amplifier to respond with the X demodulator
output in the range ±30000, full-scale being ±10000.
In floating point mode causes the lock-in amplifier to respond with the X
demodulator output in volts or amps.
Appendix E, ALPHABETICAL LISTING OF COMMANDS
Curve buffer trigger output polarity
E-25