6.7.12 Instrument Identification

ID
VER
DATE
NAME

6.7.13 Auto Default and Calibration

ADF n
6-34
requests and hence commands will be accepted from any computer on the same
network as the instrument.
Identification
Causes the lock-in amplifier to respond with the number 7230.
Report firmware version
Causes the lock-in amplifier to respond with the firmware version number, as also
displayed on the Configuration menu.
Last calibration date
The response is a six digit number giving the date of the last calibration where the
first two digits are the day, the second two digits are the month and the last two
digits are the year (in the 21
Set/Read Instrument Name
The NAME command allows the user to store an ASCII text string as a "name" to
help identify the instrument. This is especially useful when there are several
instruments connected to the same Ethernet network, since it can be used as a more
convenient alternative to the serial number as an aid to identification.
When sent with a string of up to 64 characters as a parameter, the string will be
stored. When sent without characters the instrument will respond with the stored
string. For example:
NAME Optical absorption lab lock-in amplifier
will set the name to "Optical absorption lab lock-in amplifier"; sending NAME with
no parameters will return:
Optical absorption lab lock-in amplifier
Auto Default
The ADF command performs an auto-default operation according to the following
table:
n
Significance
0
All instrument settings are returned to their factory default values
1
All instrument settings, with the exception of the communications settings, are
returned to their factory default values
2
All instrument settings, with the exception of the communications settings, are
returned to their factory default values, except the oscillator amplitude is set to
0 V rms.
NOTE: If the ADF 0 command is used when the communications settings are at
values other than their default settings, then communication will be lost.
st
century).