in the following table:
Bit
Status Byte
bit 0
command complete
bit 1
invalid command
bit 2
command parameter error
bit 3
reference unlock
bit 4
output overload
bit 5
new ADC values available
after trigger
bit 6
input overload
bit 7
data available
The LEDs are green when the corresponding bit is asserted and when this is a "no
fault" condition, and red when it is asserted but this indicates a fault. Bit 0 in each
case is on the right and bit 7 on the left.
If the mouse cursor is hovered over a bit then the bit function is indicated in a pop up
box. In Figure 5-29, for example, the cursor has been hovered over bit 3 in the Status
byte; the fact that this is red indicates a loss of reference lock, while bit 1 indicates
"command done" and bit 5 that the auxiliary ADC inputs have been triggered.
This completes the description of the Main Controls panel.

5.3.12 Oscillator: Overview

The Oscillator panel, shown above in Figure 5-30, gives access to the full set of
controls for the internal oscillator. Note, however, that the basic controls setting its
amplitude and frequency are also displayed on the Main Controls panel.
Chapter 5, WEB CONTROL PANEL OPERATION
Table 5-1, Status and Overload Byte Bit Definitions
Figure 5-30, Oscillator Panel
Overload Byte
X(1) output overload
Y(1) output overload
X(2) output overload
Y(2) output overload
CH1 output overload
CH2 output overload
CH3 output overload
CH4 output overload
5-27