Setting up your device ..................5 Navigating the touch screen .................. 7 Additional features ..................8 Connecting your AirSense 11 device and smart device ........... 9 Starting/Stopping therapy .................. 10 About the heated tubing ..................11 Therapy data ....................12 Cleaning and caring for the device ................
Page 3
Quick setup view Components 1. HumidAir™ 11 tub 2. AirSense™ 11 device 3. Power supply unit 4. ClimateLineAir™ 11 tubing 5. Mask...
AirSense 11 CPAP (including Elite) The AirSense 11 CPAP system is indicated for the treatment of obstructive sleep apnea (OSA) in patients weighing more than 66 lb (30 kg). The AirSense 11 CPAP system is intended for home and hospital use. Contraindications...
ResMed via the cloud to enable ResMed to deliver various benefits to you and your care provider(s). Additionally, some of that data may be used by ResMed (1) to comply with its legal obligations; these legal obligations include collection and analysis of device data for...
• SD card cover Notes: • The AirSense 11 device is compatible with ResMed masks. For a complete list, see the Mask/Device compatibility list on ResMed.com/downloads/devices. • The HumidAir 11 Standard tub and the HumidAir 11 Cleanable tub are the only water tubs used with the AirSense 11 device.
About your device Description Purpose Start Therapy/ Standby button Press to start/stop therapy. The LED indicator is green during standby mode, and white during therapy, Test Drive, and Mask Fit functions. Display touch screen Navigates between functions and displays information on the operating status of the device.
CAUTION Use only ResMed parts (eg, air inlet filter, power supplies), masks and accessories with the machine. Non ResMed parts may reduce the effectiveness of the treatment, may result in excess carbon dioxide rebreathing and/or damage the machine.
Page 9
Recommended masks for use with this device are listed on ResMed.com. Notes: • Do not insert any USB cable into the AirSense 11 device or attempt to plug the AC adaptor into a USB device. This may cause damage to the AirSense 11 device or USB device. •...
Navigating the touch screen The AirSense 11 device operates via a display touch screen, which allows you to access, view and change therapy and device settings. You can also track your sleep health progress. The status bar at the top of the screen may display icons at different times and may include:...
Using the touch screen: There are two actions to navigate through the touch screen: Swipe: Swipe up or down the screen to view menu options. Tap: Select a parameter setting to update. For other parameters (eg Pressure Relief, Airplane mode), tap the parameter to turn it on or tap to turn it off Prescription settings...
Test Drive feature, and tracking your sleep health progress. The app is not required to operate the AirSense 11 device. Before pairing the AirSense 11 device to a smartphone, ensure the app's latest version is installed on the smartphone. If not, download the app from the App Store ®...
Starting/Stopping therapy WARNING The machine is not intended to be operated by persons (including children) with reduced physical, sensory or mental capabilities without adequate supervision by a person responsible for the patient’s safety. To start therapy: 1. Fit your mask 2.
About the heated tubing The ClimateLineAir 11 is a heated breathing tube that delivers air to a compatible mask. When used with the device humidifier tub, ClimateLineAir 11 heated air tubing allows you to use the Climate Control feature. Note: Not all types of air tubing are available in all regions. Climate Control Climate Control is designed to make therapy more comfortable by enabling constant temperature and maintaining humidity.
Note: The temperature and humidity settings are not measured values. Therapy data The AirSense 11 device records your therapy data for viewing and adjusting by your care provider if required. The data is transferred to your care provider in the following methods:...
Page 16
To remove the SD card cover and insert SD card: 1. Push the SD card cover. 2. Remove the SD card cover and keep the SD card cover in a safe place. 3. Insert the SD card. 4. Push in the SD card until it clicks in place. To remove the SD card: 1.
Exposure to smoke, including cigarette, cigar or pipe smoke, as well as ozone or other gases, may damage the device. Damage caused by any of the foregoing, will not be covered by ResMed's limited warranty.
Disassembling 1. Hold the humidifier tub at the top and bottom, press it gently and pull it away from the device. Note: take care when handling the humidifier tub as the humidifier tub may be hot. Allow 10 minutes for the heater plate and any excess water to cool. 2.
Do not wash the air tubing in a dishwasher or washing machine. • The air filter is not washable or reusable. Checking WARNING • Discontinue use and contact your care provider or ResMed Service Center if any of the following occur: • device does not perform as usual •...
• Make sure that the humidifier tub is empty before transporting the device. Your AirSense 11 device may be taken on board as carry-on luggage. Medical devices do not count toward your carry-on luggage limit.
Troubleshooting If you have any problems, have a look at the following troubleshooting topics. If you are not able to fix the problem, contact your care provider or ResMed. Do not try to open the device. General Issues Problem/possible cause...
Page 22
Incompatible mask being used Only use equipment recommended by ResMed. Contact ResMed or see ResMed.com for more information. If you are using a nasal pillows mask with set pressure less than 7 cm O (7 hPa), SmartStop may not work and should be disabled.
3.9" (10 cm) to any part of the device. Otherwise, degradation of the performance of this equipment could result. For any serious incidents that occur in relation to this device, these should be reported to ResMed and the competent authority in your country.
4.0 cm H O (4 hPa). Flow (maximum) at set pressures The following are measured according to ISO 80601-2-70 201.12.1.103: With HumidAir 11 tub AirSense 11 and Standard AirSense 11 and Pressure AirSense 11 and SlimLine air tubing...
Page 25
Material: Blended synthetic fibers in a polypropylene carrier Efficiency: >80% (average) when tested to EN13274-7. Note: The use of a ResMed approved hypoallergenic filter will result in a small reduction in the accuracy of the delivered pressure at high leaks.
Page 26
3.94" (10 cm) separation distance. The AirSense 11 has been designed to meet EMC standards. However, should you suspect that the device performance (eg. pressure or flow) is affected by other equipment, move the device away from the possible cause of interference.
Page 27
Maximum dynamic pressure variation according to ISO 80601-2-70:2015 AirSense 11 with HumidAir 11 tub and air tubing Breath rate 10 BPM 15 BPM 20 BPM Dynamic pressure variation (cm H O [hPa]) AirSense 11 with Side cover and air tubing...
Standard: 6'6" (2.0 m) Note: The manufacturer reserves the right to change these specifications without notice. Air tubing resistance to flow and compliance information Refer to the Air tubing compliance guide in ResMed.com. Symbols Follow instructions before use. Indicates a warning or caution.
ResMed. ResMed recommends that the AirSense 11 device be inspected and serviced by an authorized ResMed Service Centre if there is any sign of wear or concern with device function. Otherwise, service and inspection of the products generally should not be required during their design life.
ResMed shall not be responsible for any incidental or consequential damages claimed to have resulted from the sale, installation or use of any ResMed product. Some regions or states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation may not apply to you.