ResMed AIRMINI User Manual
Hide thumbs Also See for AIRMINI:

Advertisement

User guide
English

Advertisement

Table of Contents
loading

Summary of Contents for ResMed AIRMINI

  • Page 1 User guide English...
  • Page 2 ResMed.com/AirMiniApp...
  • Page 3: Indications For Use

    In the US, Federal law restricts this device to sale by or on the order of a physician. Indications for use The AirMini self-adjusting system is indicated for the treatment of obstructive sleep apnea (OSA) in patients (female patients with mild to moderate OSA when using AutoSet for Her treatment mode) weighing more than 66 lb (30 kg).
  • Page 4: Adverse Effects

    • AirMini setup pack - N20: Air tubing, N20 connector, HumidX ™ HumidX Plus OR • AirMini mask pack for P10: Air tubing, AirFit P10 for AirMini mask (M), P10 cushion (S), P10 cushion (L), HumidX, HumidX Plus, AirFit P10 headgear clips.
  • Page 5 Your machine is compatible with ResMed's F20 and N20 series masks and with the AirFit P10 for AirMini. Make sure all parts and accessories used with the machine are compatible. For compatibility information, refer to www.resmed.com for more information.
  • Page 6 Your AirMini machine is designed to work with ResMed’s AirMini app. However, the AirMini app is not essential to operate the machine. For more information on the AirMini app, refer to Using your machine with the AirMini app or the AirMini app Patient eHelp.
  • Page 7 2. Connect the gray end of the AirMini tubing firmly to the air outlet. Add optional humidification (for N20 or AirFit P10 for AirMini only)
  • Page 8 2. Press the Start/Stop button or wait until the machine stops if SmartStop is activated. Comfort features Ramp Time, Pressure Relief and SmartStart are enabled on your machine by default. For more information or to change these settings, refer to the AirMini app Patient eHelp.
  • Page 9 2. On your smart device, enable Bluetooth. 3. Open the AirMini app. The first time you open the AirMini app you will be asked to consent to the use of cookies and to accept ResMed's Terms and Conditions and Privacy Policy.
  • Page 10: Cleaning And Maintenance

    Note: Once therapy starts, the indicator light will start to fade. Cleaning and maintenance It is important that you regularly clean your AirMini machine to make sure you receive optimal therapy. The following sections will help you with disassembly, cleaning and checking your machine.
  • Page 11 • The AirMini connectors and AirFit P10 for AirMini contain a vent system to prevent carbon dioxide build-up in the mask. The vent system must be kept clean and clear of contamination in order to function correctly.
  • Page 12 1. Disconnect the F20 connector from the AirMini tubing by gently twisting it and pulling it apart Disconnect the AirMini tubing (N20 and AirFitP10 for AirMini only) 1. Disconnect the vent module from the AirMini tubing by gently twisting it and pulling it apart.
  • Page 13 Unplug the machine 1. Disconnect the AirMini tubing from the machine by pressing the side buttons on the cuff and pulling it away. 2. Unplug the power supply unit from the power outlet and the machine. For mask disassembly instructions, see your mask user guide.
  • Page 14 Ultra Joy dishwashing liquid. 3. Rinse thoroughly and allow to dry out of direct sunlight and/or heat. 4. Check the AirMini tubing and replace it if there are any holes, tears or cracks. Note: Do not wash the AirMini tubing in a dishwasher or washing machine.
  • Page 15 Clean the N20 connector for AirMini 1. Clean and check the N20 connector daily. The connector should be replaced at least every six months. 2. Wash the connector in warm water using a mild liquid detergent, for example Ultra Joy dishwashing liquid.
  • Page 16 • Make sure the air filter cover is fitted at all times to prevent water and dust from entering the machine. • The use of ResMed approved hypoallergenic filter will result in a small reduction in the accuracy of the delivered pressure at high leaks.
  • Page 17: Troubleshooting

    Please check with your airline for their policy regarding medical equipment. You can use your AirMini machine on a plane as it meets the Federal Aviation Administration (FAA) requirements. Air travel compliance letters can be downloaded and printed from www.resmed.com.
  • Page 18 Problem/possible cause Solution The AirMini system may be assembled Make sure your AirMini system is correctly incorrectly. assembled. Refer to Setting up and starting therapy for more information. Air pressure in my mask seems too low / It feels like I am not getting enough air Ramp may be in progress.
  • Page 19 Problem/possible cause Solution The Bluetooth indicator light is constantly white. I cannot pair to my machine Bluetooth is disabled. Press the Bluetooth button for 3 seconds. When the Bluetooth indicator light flashes blue, the machine is ready for pairing. I cannot scan the QR code to pair my machine The camera is not focused or broken, or You can pair your machine manually by the QR code label is damaged.
  • Page 20: General Warnings And Cautions

    ResMed Service Center. • Do not open or modify the machine. There are no user serviceable parts inside. Repairs and servicing should only be performed by an authorized ResMed service agent.
  • Page 21 • The F20, N20 and P10 connectors for AirMini are fitted with vents. The F20 connector for AirMini is also fitted with an AAV.
  • Page 22: Technical Specifications

    13 ft (4 m) of X-ray or CT equipment. Never bring the device into an MR environment. CAUTION • Use only ResMed AirMini parts, masks and accessories with the machine. Non-ResMed AirMini parts may reduce the effectiveness of the treatment, may result in excess carbon dioxide rebreathing and/or damage the machine.
  • Page 23 (eg, pressure or flow) is affected by other equipment, move the machine away from the possible cause of interference. The AirMini complies with Part 15 of the FCC Rules and Industry Canada license-exempt RSS standards. Operation is subject to the following two conditions: This machine may not cause harmful interference, and this machine must accept any interference received, including interference that may cause undesired operation.
  • Page 24 The A-weighted sound power level measured according to ISO 17510:2015: Air tubing connected to AirMini F20 connector 27 dBA with uncertainty of 3 dBA with mask Air tubing connected to the AirMini N20 or P10 25 dBA with uncertainty of 3 dBA connectors with mask AirMini machine Dimensions: 5.4"...
  • Page 25 Length: 6'2" (1.9 m) Inner diameter: 0.6" (15 mm) Aircraft use ResMed confirms that the machine meets the Federal Aviation Administration (FAA) requirements (RTCA/DO-160, section 21, category M) for all phases of air travel. Wireless technology Technology used: Bluetooth Connection types:...
  • Page 26 Pneumatic flow path 1. Flow sensor 2. Blower 3. Pressure sensor 4. AAV (F20 connector only) 5. Vent 6. HumidX (N20, P10 connectors only) 7. Mask 8. Air tubing 9. Machine 10. Inlet filter General The patient is an intended operator. Pressure flow curve (ISO17510:2015) Pressure Flow [L/min]...
  • Page 27 Measurement system uncertainties In accordance with ISO 80601-2-70:2015 the measurement uncertainty of the manufacturer's test equipment is: For measures of flow ± 1.5 L/min or ± 2.7% of reading (whichever is greater) For measures of static pressure ± 0.15 cm H O (hPA) For measures of dynamic pressure ±...
  • Page 28 Flow (maximum) at set pressures The following are measured accordingly to ISO 80601-2-70:2015 at the end of the specified air tubing: Pressure AirMini machine and AirMini air tubing cm H O (hPa) L/min Notes: • The manufacturer reserves the right to change these specifications without notice.
  • Page 29: Limited Warranty

    ResMed machine please contact your ResMed office, local distributor or go to www.resmed.com/environment. Servicing The AirMini machine is intended to provide safe and reliable operation when operated in accordance with the instructions provided by ResMed. ResMed recommends that the AirMini machine be inspected and serviced by an authorized ResMed Service Centre if there is any sign of wear or concern with device function.
  • Page 30 This warranty is only available to the initial consumer. It is not transferable. If the product fails under conditions of normal use, ResMed will repair or replace, at its option, the defective product or any of its components. This Limited Warranty does not cover: a) any damage caused as a result of improper use, abuse, modification or alteration of the product;...
  • Page 31 This warranty gives you specific legal rights, and you may also have other rights which vary from region to region. For further information on your warranty rights, contact your local ResMed dealer or ResMed office. Further information If you have any questions or require additional information on how to use the machine, contact your care provider.
  • Page 32 See ResMed.com for other ResMed locations worldwide. AirMini, AutoSet, HumidX and SmartStart are trademarks and/or registered trademarks of the ResMed family of companies. For patent and other intellectual property information, see ResMed.com/ip. “Made for iPhone” means that an electronic accessory has been designed to connect specifically to iPhone and has been certified by the developer to meet Apple performance standards.

Table of Contents