When to Remove Damper and Mixer
The binary pump is equipped with a pressure pulsation damper and a static
mixer. The total delay volume of the pump is 600 – 800 μL (depending on
system pressure). The mixer has a volume of 400 μL.
For applications that require lowest delay volume (for example, fast gradient
methods or gradient applications with low flow rates), damper and mixer can
be bypassed.
Figure 9
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual
Flow path modifications of the Binary Pump
Optimizing Performance
When to Remove Damper and Mixer
4