Convert The Binary Pump To Low Delay Volume Mode - Agilent Technologies 1260 Infinity II User Manual

Binary pump
Hide thumbs Also See for 1260 Infinity II:
Table of Contents

Advertisement

4
Optimizing Performance
When to Remove Damper and Mixer

Convert the Binary Pump to Low Delay Volume Mode

The binary pump is delivered in standard configuration (damper and mixer
connected). This paragraph shows how to bypass damper and mixer and
convert the pump to low delay volume mode.
Configurations where only damper or mixer are disconnected while the other
part is still in line are not supported by Agilent Technologies.
Tools required
p/n
8710-0510
Preparations
Flush the system (water if buffers were used, otherwise isopropanol).
Turn the flow off.
1
Open the doors.
64
Description
Wrench, 1/4 - 5/16 inch, open
Wrench, 14 mm
Hex driver open, 1/4 inch
2
Use the 1/4 inch hex driver to remove fitting B from port
2 of the pressure sensor. Fold capillary end B away. It
remains unconnected.
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual