Test Functions and Calibration
Pump Elasticity Calibration

Running the Pump Elasticity Calibration

Running the Pump Elasticity Calibration from the Agilent Lab Advisor
Software
When
The initial calibration is done at the factory. It only needs to be repeated after replacement of major
pump parts (mainboard, pump drive).
Tools required
p/n
8710-0510
Parts required
p/n
G1312-67500
Preparations
Place all bottle heads in to a bottle of HPLC-grade water.
Make absolutely sure that the pump is very thoroughly flushed with the solvent to be
N O T E
calibrated before starting the procedure! Any trace of other solvents or the smallest air
bubble inside the flow path definitely will cause the calibration to fail!
If a solvent selection valve is installed flush all four solvent channels to avoid that air from
N O T E
a dry solvent intake tube is drawn into the flow path upon initialization.
1 Select the pump elasticity calibration from the test selection menu.
2 Start the test and follow the instructions.
Make sure to release the pressure by opening the purge valve when the test has
N O T E
completed. Otherwise the pump may generate an overpressure error.