1
Introduction
Instrument and Operation
Principle of Operation
The binary pump is based on a two-channel, dual-piston in-series design
which comprises all essential functions that a solvent delivery system has to
fulfill. Metering of solvent and delivery to the high-pressure side are
performed by two pump assemblies which can generate pressure up to
600 bar.
Each channel comprises a pump assembly including pump drive, pump head,
active inlet valve with replaceable cartridge, and outlet valve. The two
channels are fed into a low-volume mixing chamber which is connected via a
restriction capillary coil to a damping unit and a mixer. A pressure sensor
monitors the pump pressure. A purge valve with integrated PTFE frit is fitted
to the pump outlet for convenient priming of the pumping system.
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual
11