1
Introduction
Leak and Waste Handling
Figure 8
Infinity II Two Stack Leak Waste Concept (bench installation)
The waste tube connected to the leak pan outlet on each of the bottom
instruments guides the solvent to a suitable waste container.
Leak Sensor
Solvent incompatibility
C AU T I O N
The solvent DMF (dimethyl formamide) leads to corrosion of the leak sensor. The
material of the leak sensor, PVDF (polyvinylidene fluoride), is incompatible with DMF.
➔
Do not use DMF.
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual