Module Error Messages
Degasser: Vacuum cannot be maintained
Error ID: 8014
Pressure in degasser vacuum chamber exceeds 180 mbar limit during normal
operation.
Probable cause
1
2
3
Degasser: limit not reached
Error ID: 2244, 8053
Degasser is not ready within 8 min after startup, because the pressure inside
the vacuum chamber exceeds 120 mbar.
Probable cause
1
2
3
94
Liquid in degasser tubing.
Leak in degasser tubing or chamber.
Degasser vacuum pump defective.
Liquid in degasser tubing.
Leak in degasser tubing or chamber.
Degasser vacuum pump defect.
InfinityLab LC Series 1260 Infinity II Binary Pump User Manual
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.