Fault - Cause
-
Remedy
)iscon.lApnoea
pressure
4PNOEA PRESSURE)
'aw high
'AW fl
.pnoea volume
IPNOEA VOL)
iO2 low
:I02 Ifl
lalothane
high
IAP HAL A)
nflurane high
(AP ENF /f,
oflurane
high
'AP IS0 fi
M 8030 fault
xxx
'MB030 INOP)
Cause
Insufficient
fresh gas supply
Procure
adequate
fresh gas supply
Mean pressure
Pmean
smaller than -2 mbar
Airway pressure
Paw
smaller than -7 mbar
Insufficient
fresh gas supply
Procure
adequate
fresh gas supply
Breathing
or ventilation
has stopped
Leak in hose system
Check
hose system
Upper limit value for airway
pressure
has been exceeded,
Kink in ventilation
hose, stenosis
Check
hose system on
anaesthetic
machine
Pressure
limit set too high
Adjust pressure
limit on
anaesthetic
machine
Overpressure
valve switched
off
Switch
on pressure limit on
anaesthetic
machine
Breathing
or ventilation
has stopped.
No volume expired during
15 seconds
Insufficient
fresh gas
supply
Tube blocked/kinked
Leak in hose system
Check ventilator
Procure
adequate
fresh gas supply
Check
hose system
The inspiratory
02 concentration
is below the lower limit value
Check
02 supply. check
setting of flowmeter
on blender
The appropriate
anaesthetic
agent
concentration
in the fresh gas is
higher than the upper limit value
Check setting on anaesthetic
vaporiser
Incorrect
anaesthetic
agent has
been selected
Internal hardware
fault
with error number
Machine is not ready
for
operation
Call DrlgerService,
give error number
* Messages
in parenthesis
are displayed,
or printed on the peripheral
machines
PM 8020 and PM
8010.
Or on a
PC or on a printer.