Operation
Limit values - reduced monitoring
For operating
conditions
such as manual ventilation
or
spontaneous
breathing,
monitoring
can be reduced
to
the parameters
A-Vap with upper limit value
02 with lower limit value
Paw with upper limit value
thus preventing
alarms which are not required.
Press z-limit values<
button.
Display:
Limit values <IPPV>
Mode 02
VE
Paw A-Vap
Select >Mode<
and confirm.
Display:
MODE
<IPPV>
MANISP
IPPV
default
Select >MANISPc
and confirm.
Display:
limit values
MANISP
active
=
reduced
monitoring
Activating
all limit values
Set, as described
above,
but
3
Select >IPPV<
and confirm.
2
Display:
Limit values
IPPV active
All limit values are effective
at the set values
Selecting
default limit values
Set. as described
above,
but
3
Select >&fault<
and confirm.
2
Display:
Limit values
default active
These default limit values are set by the manufacturer
and
apply after each restart
(table. page 19).
22
33