Dräger PM 8030 Instructions For Use Manual page 45

Table of Contents

Advertisement

Technical
Data
Inspiration gas temperature
Range
Resolution
Accuracy
Anaesthetic agent concentration
Range
Resolution
Accuracy
Data interface
(not for the control of the
anaesthetic or intensive
care ventilator)
RS 232-C
numerical display, VF display, 2 digit
20 to 50°C
1°C
IO.5;C
in measuring range 30 to 41°C
numerical 7-segment display, 2 digit
0 to 9.9% by vol.
0.1% by vol.
5 10 % of measured value, at least > 0.1% by vol.
Plug
Pin plates
Galvanic separation
Operating
Data:
Voltage
Current
Mains fuse
Dimensions W x H x D
Weight
Classification
25 pole Sub 0
1 screening
2 - TxD
3. RxD
7.GND
1.5 kV
100 to 240 V 5OkiO Hz
max. 0.4 A
T 2 A DIN 41662 (2x)
425 x 66 x 300 mm
3.3 kg
.
I, type 6 R
co n orming to DIN IEC 601/l
para. 19, table 4
f
max. permissible earth leakage current 0.5 mA
Temperature and flow sensor. galvanic, separated from
protective conductor type BF
permissible patient current 0.1 mA
Manufacturer's
certificate
for radio suppression
The manufacturers certify that PM 6030 is radbsuppressed
to conform to the guidelines contained in Regulations
1046 of the German Federal Post.