02 Measurement
Measuring
principle and signal processing
The 02 sensor functions
according
to the galvanic
cell principle.
Oxygen
molecules
from the gas mixiure for measurement
diffuse through
a plastic
membrane
into an electro-chemical
cell and are reduced
at noble metal
electrodes.
Simultaneously
oxidation
occurs at a base metal electrode
which is eroded as
a result of the oxidation
process,
and thus the life of the sensor is limited. The
current flowing
through
the cell is proportional
to the oxygen partial pressure
in the gas mixture.
At a constant
pressure
and temperature
of the gas mixture being measured,
the measured
value is directly proportional
to the oxygen
partial pressure.
Particularly
characteristic
of the 02 sensor are the two electrically
separated
cathodes
which give two independent
redundant
measuring
signals. Both
measuring
signals are evaluated
electronically
and the mean value of both
individual signals is displayed.
Where the individual
signals differ beyond
a permissible
tolerance
due to
external or internal malfunction,
the display will switch off automatically.
This
prevents faulty displays and subsequent
faulty interpretation.
I
Anaesthetic
agent measurement
Measuring
principle
The
anaesthetic
agent measurement
works with the Drsger
Iris sensor (IRIS
= InfraRed Inhalation
Sensor).
The anaesthetic
agents
halothane,
enflurane
and isoflurane
absorb a pulsed light beam inside the measuring
chamber.
At
a constant
pressure,
the reduction
of light intensity
(compared
to a calibration
point) can be described
approximately
as the sum of two exponential
functions.
In the selected
wavelength
ranges changes
in the composition
of the gas
mixture (NzO etc.) give no measurable
cross sensitivity
at the required
accuracy.
PM 8030 is calibrated
by the manufacturer
at a temperature
of 22°C. normal
pressure of
1013
hPa.
Since measurement
depends
on gas density, the measured
value must be
corrected
when there is any significant
deviation
from normal pressure
(1013
hPa) and temperature
(22%):
Ckorr. =
c
It?.
273+T
P
295
:a,,.
=correct
measured
value
= measured
value
PO
= normal pressure
10 13 hPa
Y
= actual atmospheric
pressure
hPa
= actual temperature
"C
No correction
for pressure
is required
for comparing
the concentration
set on
the Drsger anaesthetic
agent vaporiser
Vapor 19.n and the display value on
PM 8030.
See Driger
Vapor 19.n Instructions
for Use.