Dräger PM 8030 Instructions For Use Manual page 36

Table of Contents

Advertisement

Checking Function
after assembly. at least every 2 weeks,
Self-test
of machine
l
Switch on mains power:
The monitor carries out a self-test:
The internal programme memory is tested.
The software and language versions are displayed.
All LEDs and display elements are lit; a single tone
sounds.
Display:
Dr8ger
PM 8030
The alarms for monitoring are tested.
The self-test is completed in about 30 seconds.
For operation with Monitorbus:
0 Set ventilator to IPPV.
1 The green LED goes out, the PM 8030 is in
measuring mode.
0
Calibrate flow sensor, page 13.
0 Calibrate 02 sensor, page 11.
0 Calibrate anaesthetic agent sensor, page 15.
l
Check pressure measurement, page 17.
l
Check temperature measurement, see below.
Checking
temperature
measurement
Pmean
AW-Temp
Freq
xx
Remove temperature sen.wr from Y-piece and expose
to ambient air.
Select the display for ailway temperature:
Aftar about 1 minute, room temperature is displayed.
Re-insert temperature sensor in Y-piece.
36