Operation
Setting 02 limit value
Recommendation:
For 02 concentrations
lower than 50% by vol.:
lower limit value 5% by vol. lower than the actual 02
concentration.
Where air is applied post-operatively:
set lower limit value to 20% by vol.
For 02 concentrations
greater than 50% by vol.:
lower limit value 10 % by vol. lower than the actual 02
concentration.
~Depending on clinical indications, set or switch off upper
limit value.
Setting Paw limit value
Set lower limit value about 8 mbar below plateau
pressure.
Where there is no pleateau pressure, set at about 8 mbar
below peak pressure.
The lower limit value may be switched off.
Set the upper limit value about 10 mbar above peak
pfeSWie.
Setting minute volume (GE) limit values
Recommendation:
Set lower limit value 20% lower than minute volume
Set upper limit value about 20% higher than minute
volume
20