Operation
Carrying out function check
Immediately
before daily use as described
in the checklist
on the anaesthetic
machine
Monitor
self-test
Switch
on mains power supply at the back.
The monitor
carries out a self-test:
The internal programme
memories
are tested.
The software
and language
versions
are displayed.
All LEDs and display elements
are lit; a single tone
sounds.
Display:
Drtiger PM 8030
The alarms for monitoring
are tested.
The self-test
is completed
after about 30 seconds.
For operation
with Monitorbus
0
Switch
ventilator
to IPPV:
1
The green LED goes out, PM 8030 is in measuring
mode.
Checking
02
measurement
2
0
3
l
Press >Standby<
button, green LED goes out =
measuring
mode
Remove
02 sensor from cap and expose to ambient
air for at least 2 minutes.
Display:
21 vol.
%
02
Replace
02 sensor on cap.
16