Preparation
Connecting
flow sensor and pressure
measuring line
with measuring connection
M 28 833
0
1
2
1
3
Screw flow sensor into housing.
Unscrew expiratory valve.
Screw the measuring connection on with the flow
sensor in place.
Screw expiratory valve back.
Push plug of pressure measuring line into coupling -
as far as it will go. Tilt hose upwards so that
condensate can flow downwards.
Connecting temperature
sensor
Adapter and accessories:
Y-piece
M 30 543
Hose clip (set of 10)
84 04 047
Push temperature sensor into hole on the Y-piece as
far as it will go.
After replacing the Y-piece, align sensor vertically so
that condensate will not be able to get into it.
Attach sensor cable to inspiratoty hose with hose
clips.
A3
Connecting other machines
via
RS 232
C-Interface
.
PM 8010 Patient Monitor
with data cable "PM 8010"
85 00 337
.
PM 8020 Data Manager
with data cable "PM 8020"
Printer e.g.:
Desk Jet Printer (Hewlett Packard)
Think Jet Printer (Hewlett Packard)
PA 565 (NEC)
with data cable
0 Connect and secure plug to both machines.