Configuration
Interface
protocol for printer
PM 8030
is
preset by the manufacturer
for the
connection
of a printer:
Transfer
rate
9600 baud
interval
time for measured
value
5 minutes
Output
of warning
messages
"Warning"
on
No output of caution
messages
"Caution"
off
1
2
3
2
4
2
4
2
4
3
2
4
2
4
2
4
3
2
Switch
PM 8030 to Standby;
green LED is lit.
Display:
Standby
Keep button pressed
for at least 3 seconds:
Display:
configuration
m
RS 232 sensor
Select
>RS 232~ = Urn
knob and
confirm
= press knob
Display
RS
232
<printer>
MEDIBUS
printer rate
Select xate<
and confirm.
Display:
rate/100
baud
<96>
0 12 24 48 96 192
The
actual setting is indicated
by < > on the right in
the top line.
Select desired
baudrate
and confirm.
Press button briefly.
Display
RS 232
<printer>
MEDIBUS
printer rate
Select
sprinter<
and confirm.
Display:
printer
<5>
&Q
warning caution
Selecr
>data<
and confirm.
Display:
data/min
<5>
1 2 5 10 on
Select
desired
interval time for measured
value print-
out and confirm.
Press
button
and select "Warning".
Display:
warning
<on>
on
off
Warning <on> =
warning
messages
are given
Warning
<off> = warning
messages
are not given.