Fault
-
Cause
-
Remedy
lessage
:aution
messages
lin. vol. low
IMV Jf)
alothane
low
'AP HAL J/j
nflurane low
'AP ENF if)
oflurane
low
'AP IS0 J/-)
Cause
Lower limit value for minute volume
has been crossed
Tube blocked/kinked
Leak in breathing
system
Loss of volume through
pressure
limitation
Lung compliance
has been reduced
Flow sensor not calibrated,
or faulty
The appropriate
anaesthetic
agent
concentration
in the fresh gas
is lower than the lower limit value
Incorrect
anaesthetic
agent
has been selected
Remedy
Check
tube
Seal breathing
system.
Correct
ventilation
parameters
Calibrate
flow sensor,
page 13; replace
if required
Check
setting and filling
level of anaesthetic
vaporiser
02 high
'102 fT)
Oz.flush used, the inspiratory
02 concentration
is above the
upper limit value
Check
02 setting on the
flowmeter
block or on the blender
in. vol. high
.MV /r)
ul/-Temp. high
W-Temp /'T)
Upper limit value for the minute
volume has been exceeded
Flow sewor
not calibrated
or faulty
The inspiratory
breathing
gas
temperature
is above 40°C
Calibrate
sensor,
page 13
replace
if required
Switch
off humidifier.
When temperature
has dropped
to
about 37°C set low heating level
neck RS 232
Fault in data connection
or cable
Check
cable
Fault in connected
machine
Check
machine