Dräger PM 8030 Instructions For Use Manual page 15

Table of Contents

Advertisement

Calibration
Calibrating anaesthetic
agent sensor
when sensor has been replaced
every
24 hours
Allow
3
minutes warming-up
time after switching
on the
machine.
On anaesthetic
machine
l
Set anaesthetic
vaporiser to 0 (zero point stop
button engaged),
otherwise
faulty calibration
will occur.
Set 02
flow of about 10 Llmin on flowmeter block or
on blender.
l
Flush anaesthetic agent seneor for at least 10
seconds.
Do not use 02 flush for flushing.
After
switching
on and after request
by machine
Or
1
2
3
2
Press calibration button and select A-Vap sensor.
Display:
Calib. A-Vap sensor?
Ye
II0
Select >yes<
and confirm.
Display:
A-Vap sensor
flushed with Oz/air?
Continue with anaesthetic vaporiser at 0 and 02 flow of
about 10 Umin.
3 Confirm = press knob.
2 Display:
A-Vap sensor
calib.
4 A-Vap display
CAL
Calibration is completed in about 0 seconds
2 Display:
A-Vap sensor
calib. completed
If there is a faulty sensor
or
cable:
2
Display:
A-Vap sensor
not
ready to measure
15
32