Dräger PM 8030 Instructions For Use Manual page 33

Table of Contents

Advertisement

Care
Stripping-down
.
Detach temperature
sensor from Y-piece;
detach Y-
piece from ventilation
hoses; remove cable from hose
clips.
.
Disconnect
plug at the back; do not pull on the cable.
l Unscrew
the housing and remove flow sensor
When contamination
is likely:
l Remove flow sensor = disposable
or
0
Reuse
flow sensor as long as it is possible
to
calibrate
it.
The sensor cannot be autoclaved
or disinfected.
Loosen
plug on pressure
measuring
line = push ring
on coupling
back.
Remove pressure
measuring
line and filter from the
back of the machine and drain condensate
from the
measuring
line.
0
Remove 02 sensor and disconnect
plug from the
back of the machine;
do not pull on the cable.
33
C6