Fault - Cause
-
Remedy
02
sensor
not
ready
IO measure
(FiOz INOP)
Flow sensor not
ready to measure
(FLOW INOP)
A-Vap sensor not
fitted?
(A-VAP INOP)
A-Vap sensor
warming up time
(A-VAP INOP)
I---
A-Vap sensor not
ready to measure
(A-vAP INOP)
AW.Temp. Sensor not
ready to measure
(AW-TEMP INOP)
Settings lost
Cause
Sensor caii'xated incorrectly
Sensor has been replaced but not
calibrated
Sensor's !iie exceeded
Faulty sensor cable
Sensor caiibrated incorrectly
Faulty sensor
Sensor repiaced but not calibrated
Faulty cabie
Faulty sensor
Sensor
plug
not connected
Sensor is siill warming up
Sensor
caliorated incorrectly
Sensor or cable not connected,
or faulty
Internal banFry for data storage is
discharged
Temporary -voltage failure
Internal. au:omatic new start
Remedy
Calibrate sensor, page 11
Calibrate sensor
Replace sensor capsule,
page 7, and calibrate
Change housing of 02
tensor
Calibrate sensor, page 13
Replace sensor and calibrate,
page 13
Calibrate sensor
Replace cable and calibrate sensor
Confirm = press knob, the
advisory message disappears
Call DrigerService
Connect plug at the back, confirm
Await 3 minute warming-up time
Calibrate sensor, page 15
Connect sensor or use new sensor
Calibrate all sensors. the machine
is then ready for operation: Call
Dr~gerService without delay
Monitoring is reactivated after a
few seconds