3.1.2 Cleaning and Disinfecting of the Unit

1. Open the lid, before turning off the unit. Disconnect from the power supply.
2. Open the rotor nut, by turning the rotor key counter-clockwise.
3. Remove the rotor.
4. For cleaning and disinfection of the unit and the rotor chamber, use the above mentioned
cleaner.
5. Clean all accessible areas of the device and its accessories, including the power cord, with a
damp cloth.
6. Wash the rubber seals and rotor chamber thoroughly, with water.
7. Rub the dry rubber seals with glycerol or talc, to prevent these from becoming brittle. Other
components of the unit, e.g. the lid lock, motor shaft and rotor, should not be greased.
8. Dry the motor shaft with a soft, dry and lint-free cloth.
9. Examine the unit and accessories for damage.
Remove adherent dust, at least every 6 months, from the ventilation slots in the centrifuge, by using
a soft brush.
*Before doing so, please switch off the unit and disconnect from the power supply.

3.1.3 Cleaning and Disinfecting of the Rotor

1. Clean and disinfect: the rotors, rotor lids and adapters, with the cleaner previously mentioned
above.
2. Use a bottle brush to clean and disinfect the rotor bores.
3. Rinse the rotors, rotor lid and adapter, with clear water. Particularly, the drillings of the angle
rotors.
4. When drying the rotors and accessories, set on a towel. Place the angle rotors, with bores
down, to dry.
5. Dry the rotor cone with a soft, dry and lint-free cloth, check for damage. Do not grease the rotor
cone.

3.1.4 Disinfection of Aluminum Rotors

In case of infectious material spilling into the centrifuge, the rotor and rotor chamber have to be
disinfected, promptly after the run. Rotors may be autoclaved, at a maximum temperature of
121°C/250°F.

3.1.5 Disinfection of PP-Rotors

Autoclaving
he recommended time for autoclaving: 15 – 20 min at 121°C/250°F, (1 bar)
ATTENTION: The sterilization time of 20 min. must not be exceeded. Continuous sterilization
will cause reduction in the mechanical resistance, of the plastic material.
Before autoclaving the PP-rotor and adapter, thoroughly clean to avoid the burning of dirty residue.
Please disregard any consequences of chemical residues to plastic materials, at ambient
temperatures. At high temperatures, autoclaving residue may corrode and destroy the plastic. The
objects must be thoroughly washed with distilled water, after the cleaning, but before the
autoclaving. Residues of any cleaning liquids, may cause fissures, whitening and stains.
© Hermle Labortechnik GmbH
Z446_V2.21_eng
TROUBLE SHOOTING