Table 6-4: General Purpose Rf Measurement Characteristics; Table 6-5: Amplitude Vs. Time Measurement Characteristics - Tektronix H600 RFHawk Service Manual

Spectrum analyzer
Table of Contents

Advertisement

Specifications
6–6

Table 6-4: General purpose RF measurement characteristics

Characteristic
Description
General Purpose RF Channel Power Measurement
Measurement
1 kHz – 20 MHz
Bandwidth
Range
Accuracy
≤1.2 dB; +20 dBm to –60 dBm; Resolution BW < 100 kHz
+20 dBm to –40 dBm; Resolution BW ≥ 100 kHz
1 MHz to 3.2 GHz , preamp off
≤2 .4 dB; –60 dBm to –75 dBm; Resolution BW < 100 kHz
–40 dBm to –55 dBm; Resolution BW ≥ 100 kHz
10 MHz to 3.2 GHz , preamp on
≤1.8 dB; +20 dBm to –50 dBm; Resolution BW < 100 kHz
+20 dBm to –40 dBm; Resolution BW ≥ 100 kHz
3.2 GHz to 6.2 GHz , preamp off
≤3 dB; –50 dBm to –75 dBm; Resolution BW < 100 kHz
–40 dBm to –55 dBm; Resolution BW ≥ 100 kHz
3.2 GHz to 6.2 GHz , preamp on
Specifications apply for default control settings (Auto RBW, Auto Level)
Occupied Bandwidth Measurement
Percent
50% - 100% in 1% steps
Power
Inclusion
Range
RF Field Strength
Channel
Same as Channel Power
Bandwidth
Range
Accuracy
Same as Channel Power

Table 6-5: Amplitude vs. Time measurement characteristics

Characteristic
Description
Center Frequency
Range
Minimum center frequency = 10 kHz + Span / 2, preamp off
Minimum center frequency = 10 MHz, preamp on
Maximum center frequency = 6.2 GHz
Resolution
1 Hz
Span
Range
5 kHz to 20 MHz in 1-2-5 sequence
Sample rate is automatically set based on the span selection