Performance Verification

Procedures

Diagnostic Tests
1–6
A test record table is provided. (See page 1-4, Test Record.) You can photocopy
the page and record your test results in the table.
This section contains procedures to verify that the instrument platform is
functioning properly.
1. Select Start > Programs > Tektronix Utilities > Diagnostics.
2. Tap the Diagnostics tab.
3. To view the test selections, tap the + symbol in front of All Modules, All
Tests. Then tap the + symbol in front of Platform.
4. Select the test(s) to run. The default selection runs all tests. However,
some diagnostic tests require special equipment or custom fixtures (such as
a DB-9F serial loopback connector, PCMCIA ATA flash memory cards,
PS/2-compatible keyboard, Microphone, USB Flash Memory Device, and a
network interface connected to a DHCP-enabled network server). Clear the
box next to any test that requires equipment or custom fixtures that you do not
have available for the tests, otherwise these tests will indicate that they fail.
5. Select one of the test run options:
Loop. Use the up/down arrows to set how many times to repeat the test(s).
Until Fail. Run the selected test(s) until a failure occurs or the user selects
Stop.
Continuous. Run the selected test(s) until the user selects Stop.
6. Tap Run to begin the tests.
H600 & SA2600 Service Manual