Table 1-2: Test Voltages And Limits - Tektronix H600 RFHawk Service Manual

Spectrum analyzer
Table of Contents

Advertisement

USB Flash Memory Device
SPI Test
SMBus Test
Flash File System Test
Power Test
H600 & SA2600 Service Manual
Tests the USB port using a Flash Memory drive.
Test
NOTE.
This test requires a USB Flash Memory Device inserted into the USB slot.
If a Flash Memory Device is not installed when the test is initiated, the user is
prompted to insert a Flash Memory Drive or fail the test. The test consists of
the following:
Creating a data file
Writing a fixed size pattern to the file
Closing the file
If an error occurs during any portion of the test, the test terminates and reports a
Test failed in the Failure Log. A Test failed is also recorded if a Flash Memory
Drive is not installed when prompted.
Check the external connectivity and function by setting the instrument module
interface port to loop back mode and sending, receiving, and validating device
selects and frequency changes.
No user interaction required. Select Run.
Verify communications between the host and power management processor.
No user interaction required. Select Run.
Check the read/write function and data integrity of the internal flash file system.
The test uses standard API functions to create a file, write a data pattern to the
file, read the file, and validate the data.
Verify the instrument platform processor board and application module voltages
are within their nominal range. See Table 3-2 for voltage limits.
NOTE.
The Power Test does not check battery voltages.
This test uses internal voltage sensing. No user interaction required. Select Run.

Table 1-2: Test voltages and limits

Source
Processor board
Processor board
Voltage (VDC)
+1.5
+3.3
Performance Verification
Limits (VDC)
10%
10%
1–11

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sa2600

Table of Contents