Table 6-1: General Performance Characteristics; Specifications - Tektronix H600 RFHawk Service Manual

Spectrum analyzer
Table of Contents

Advertisement

Specifications
H600 & SA2600 Service Manual
This section lists the electrical, environmental, and physical specifications for the
H600 and SA2600 Spectrum Analyzer. All specifications are guaranteed unless
labeled "Typical." Typical specifications are provided for your convenience and
are not guaranteed.

Table 6-1: General performance characteristics

Characteristic
Description
RF Input
Operating
10 kHz – 6.2 GHz
Frequency
Operating Range for some features may be restricted. If so, the
Range
restricted range is defined with the relevant feature.
Maximum
+20 dBm peak envelope power
Operating
This is the maximum input level at which the instrument will meet its
Input Level
performance specifications.
For a signal without any amplitude variation, peak envelope power
= RMS.
Maximum
50 W rms below 3.2 GHz
Input Power
15 W rms between 3.2 GHz and 6.2 GHz
without
Damage
IF Output
Output
50 Ω
Impedance
IF Center
140 MHz, non-inverted
Frequency
IF 3 dB
24 MHz nominal
Bandwidth
IF Output Level
–12 dBm @ 1.0 GHz
–12 dBm @ 1.2 GHz
–10 dBm @ 1.6 GHz
–11 dBm @ 4.35 GHz
–16 dBm @ 5.0 GHz
–22 dBm @ 5.75 GHz
0 dBm CW signal applied to the input, instrument reference level set to
0 dBm, room temperature.
IF Output Level
± 3.5 dB for a CW input signal, 0 dBm input, 1 GHz input frequency,
Accuracy, Typical
0 dBm reference level, room temperature
IF Output
± 1.5 dB up to ± 3 MHz from center frequency
Flatness, Typical
± 3 dB up to ± 10 MHz from center frequency
Internal Timebase
Error, Factory
± 0.5 PPM from 0 °C to 50 °C
Calibration
± 1.0 PPM aging/year
Corrected
Twenty minute warm-up period required to meet accuracy specification