Table 6-1: General performance characteristics (cont.)
Characteristic
Description
Trigger
Range: –160 dBm to +20 dBm
Threshold
Resolution: 1 dB
Trigger
5 kHz to 20 MHz in 1-2-5 sequence
Bandwidth
10 MHz < Span ≤ 20 MHz: trigger BW = 20 MHz
Range
5 MHz < Span ≤ 10 MHz: trigger BW = 10 MHz
2 MHz < Span ≤ 5 MHz: trigger BW = 5 MHz
1 MHz < Span ≤ 2 MHz: trigger BW = 2 MHz
500 kHz < Span ≤ 1 MHz: trigger BW = 1 MHz
200 kHz < Span ≤ 500 kHz: trigger BW = 500 kHz
100 kHz < Span ≤ 200 kHz: trigger BW = 200 kHz
50 kHz < Span ≤ 100 kHz: trigger BW = 100 kHz
20 kHz < Span ≤ 50 kHz: trigger BW = 50 kHz
10 kHz < Span ≤ 20 kHz: trigger BW = 20 kHz
5 kHz < Span ≤ 10 kHz: trigger BW = 10 kHz
2.5 kHz < Span ≤ 5 kHz: trigger BW = 5 kHz
Spectrum mode only: Span ≤ 2.5 kHz: trigger BW = 2.5 kHz

Table 6-2: Spectrum analyzer characteristics

Description
Characteristic
Center Frequency
Range
10 kHz to 6.2 GHz, preamp off
10 MHz to 6.2 GHz, preamp on
Setting
1 Hz
Resolution
Span
Range
1 kHz to 6.2 GHz
Setting
1 Hz
Resolution
Resolution Bandwidth
Range
10 Hz to 3 MHz (Manual RBW)
10 Hz to 1 MHz (Auto RBW)
Setting
1 Hz
Resolution
Spectral Purity
Displayed
–153 dBm, 10 MHz to 2 GHz, 10 Hz RBW
Average
–152 dBm, 2 GHz to 4 GHz, 10 Hz RBW
Noise Level,
–151 dBm, 4 to 5 GHz, 10 Hz RBW
Preamp On
–145 dBm, 5 to 6.2 GHz, 10 Hz RBW
Reference Level ≤ (DANL + 90 dB)
Specifications