Technical description
2.5
Technical data
General comments
Measuring ranges
4, can be switched over internally or
externally;
manual and automatic measuring range
changeover possible
Autoranging
Hysteresis, selectable
Smallest possible
0 - 10 ppm
measuring span
Threshold of detection
0.1 ppm
Largest possible
99,999 ppm
measuring span
Concentration units
C
, C
1
(ppm)
Measured-value
Digital concentration reading
display
(5 positions with floating point)
Resolution of the digital
0.1% of the measured value
display
Position of normal use
Front panel vertical
Conformity
CE symbols to EN 50081-1, EN 50082-2
Oven temperature
Adjustable between 100-200 °C
(depending on application)
Design, housing
Degree of protection
IP 20 as per EN 60529
Dimensions
See "Dimensional drawings" diagram
Weight
approx. 23 kg
Electrical features
Auxiliary energy
100 - 120 V AC (nominal consumption
range 90 -132 V), 48 - 63 Hz, or
200 - 240 V AC (nominal consumption
range 180 - 264 V), 48 - 63 Hz
Power consumption
approx. 150 VA in operation
approx. 350 VA during warming-up
phase
EMC – resistance to
Complies with standard requirements of
interference
NAMUR NE21 (August 1998)
(electromagnetic
compatibility)
Electrical safety
To EN 61010-1,
overvoltage category II
Fuse-protection values
100... 120 V: 4.0 T/ 250
200... 240 V: 2.5 T/ 250
2-14
3
, C
, mg C/m
3
6
Gas entry conditions
Permitted measuring-
gas pressure
-
With pump
Atmospheric pressure
installed
Measuring-gas flow rate
18 ... 60 l/h (0.3 to 1 l/min)
Measuring-gas
0 - 200°C
temperature
Measuring-gas
< 90% RH
humidity
(RH = relative humidity)
Consumption figures for the gases
Combustion air
approx. 350 ml/min
Hydrogen
approx. 20 ml/min
Sample
approx. 1 l/min
Neutral gas /
approx. 2 l/min
calibration gas
Time behavior
Warming-up time
At room temperature, approx. 2 - 3 h
Display delay (T90)
2 - 3 s
Attenuation (electrical
0 -100 s, parametrizable
time constant)
Dead time (time for
With filter 2 - 3 s
flushing out gas path
inside unit at 1 l/min)
Time for signal
< 1 s
p
rocessing within unit
Measurement behavior
Output signal fluctuations
< 0.75% of smallest measuring
range possible according to
nameplate with electronic
attenuation constant of 1 s (this
corresponds to +/- 0.25% at 2 σ)
Zero-point drift
< 0.5% per month of smallest
measuring range possible according
t
o nameplate
Measured-value drift
< 1% per week of measuring span
concerned
R epeatability
< 1% of measuring span concerned
Linearity deviation
< 1% of measuring span concerned
Operating manual – A5E00222135-01
FIDAMAT 6 gas analyzer